8.5 Sample Index PCR
-
Choose the appropriate sample index sets to ensure that no sample indices overlap in a multiplexed sequencing run. Record the 10x sample index name (PN-3000431 Dual Index Plate TT Set A well ID) used.
-
Add 50 μl Library Amp Mix to 30 µl sample.
-
Add 20 μl of an individual Dual Index TT Set A to each well and record the well ID used. Pipette mix 5x (pipette set to 90 μl). Centrifuge briefly.
-
Incubate in a thermal cycler with the following protocol.
Lid Temperature
Reaction Volume
Run Time
hh:mm:ss
105°C
100 µl
~25-40 min
Step
Temperature
Time
1
98°C
00:00:45
2
98°C
00:00:20
3
54°C
00:00:30
4
72°C
00:00:20
5
Go to step 2, see below for # of cycles
6
72°C
00:01:00
7
4°C
Hold
The total cycles should be optimized based on 25% carry forward cDNA yield/input calculated during cDNA QC & Quantification
Recommended cycle numbers cDNA Input
Total Cycles
0.25-50 ng
14-16
50-250 ng
12-14
250-600 ng
10-12
600-1,100 ng
8-10
1,100-1,500 ng
6-8
>1500 ng
5
-
Store at 4°C for up to 72 h or proceed to the next step.