3.2
Load GEM-X Chip
|
GEM-X MO Chip, gasket attached Representative chip image
Pipette technique
|
|
a. If loading less than 8 samples/chip, add 50% glycerol solution to each unused well in row 1, 2, and 3
DO NOT add 50% glycerol solution to the wells in top row labeled NO FILL. DO NOT use any substitute for 50% glycerol solution. |
Glycerol
|
|
b. Prepare Gel Beads
|
Prep Gel Beads
|
|
|
|
|
d. Load Row Labeled 1
|
60 μl Master Mix + Cell Suspension in GEM-X MO Chip
|
|
e. Load Row Labeled 2
|
60 μl Gel Beads
|
|
f. Load Row Labeled 3
|
250 μl Partitioning Oil B
|
|
g. Prepare for Run
Run the chip in Chromium X series instrument (X/iX) immediately after loading the Partitioning Oil B. |
GEM-X MO Chip, closed
|